Before operation, make sure User’s Manual

and maintenance schedules are observed. X-ray must only be operated by trained

personnel in a controlled setting otherwise X-ray may be dangerous to patient,

operator or ancillary people.

 

 

 

 

 

Pregnant women are generally advised to

avoid X-rays to minimize the chances of causing any harm to the fetus unless

necessary.

 

 

 

 

Personal

exposure monitoring device and radiation protection equipment must be provided

to operators. Maximum permissible whole-body or partial body radiation dose of

operator in accordance with local regulations shall be established. The record

shall be checked before each operation.

 

 

 

 

 

The XR-01 Dental Portable X-Ray System is

intended to be used by trained dentists and dental technicians as an extraoral

x-ray source for producing diagnostic x-ray images using intraoral image

receptors. Its use is intended for both adult and child subjects

 

 

 

 

 

Rolence Dental Portable X-Ray system XR-01

is designed for use in radiography to find hidden dental structures, malignant

or benign masses, bone loss, cavities and other areas of dental diagnosis. The

controlled x-ray beam penetrates oral structures at different levels, which

depending on varying anatomical densities, and then exposures onto image

receiver such as film or Dental Digital Sensor to form a radiography image.

 

 

 

 

 

The owner/operator is responsible for

verifying continued compliance exposure rates, leakage radiation, alignment of

the useful beam, and the calibration of kVp and mAs. Annual verification by a

qualified service technician may be required by federal law. Compliance with

applicable statutory and regulatory requirements is the responsibility of the

owner/operator. Consult local, state, and/or federal agencies regarding

specific requirements and regulations applicable to the use of this type of

medical electronic equipment.

 

 

 

 

 

Ensure the adaptor of Dental Portable X-Ray

is unplugged before attempting to clean. To make sure that power is off while

cleaning. Use a non-alcohol based disinfectant only - wipes or a cloth dampened

with liquid or spray. The system is not designed to be subjected to any kind of

sterilization procedure and not designed to be used to sterilize anything else.

 

 

 

 

 

Observe the following safety procedures to

avoid electric shock or serious injury to operators and patients and to avoid

system malfunction.

 

 

● Under no circumstance should the safety

interlocks in the system be bypassed, jumped, or otherwise disabled.

 

 

● Under no circumstance should the system

or assembly covers be removed by anyone other than a Rolence trained service

representative.

 

 

● Do not place food or beverage containers

on any part of the equipment. They can trip over and introduce conductive

substances into the electrical circuitry.

 

 

 

 

 

NOTE: The system of Rolence Dental Portable

X-Ray is not waterproof. Water, soap, or other liquids, if allowed to drip into

the equipment, can cause electrical short circuits leading to electric shock

and fire hazards. If liquids should accidentally spill into the system

electronics, do not connect the power cord to a supply connection or turn the

system on until the liquids have dried or evaporated completely.

 

 

 

 

 

The user should develop an emergency

procedure for the area in which the system is used that includes the following

safety measures:

 

 

● Turn the system off and unplug the main

power cable from the outlet.

 

 

● Evacuate everyone from the area.

 

 

● Call for help.

 

 

● Use only a fire extinguisher of a type

approved for electrical fires.

 

 

 

 

 

NOTE: The use of the wrong type of fire

extinguisher presents electrical shock and burn hazards. To avoid these

hazards, a fire extinguisher which has been approved by the appropriate local,

state, and federal codes must be available in the room where Dental Portable

X-Ray is being used.

 

 

 

 

 

This Portable X-Ray tube assembly produces

ionizing radiation when energized. Hence, when, used, it can be better that the

authorized operators wear X-ray Protective devices like X-Ray protective apron,

glove or thyroid cover, etc.

 

 

It is imperative that the owner designates

areas suitable for safe operation and service of the system and operator ensure

that it is used only in these designated areas.

 

 

It is the responsibility of the owner to

ensure that all personnel wear protective clothing and radiation monitoring

devices while using this system.

 

 

 

 

 

According to FDA guidance “Radiation Safety

Considerations for X-Ray Equipment Designed for Hand-Held Use” and IEC

60601-1-3 standard, leakage radiation shall not exceed 0.88 mGy/hour and 0.25

mGy/hour respectively. All shipped Dental Portable X-Ray XR-01 have been

inspected strictly with leakage limit 0.04 mGy/hour, which is 6.25 times than

IEC regulations.

 

 

 

 

 

 

 

The backscatter shield is designed to

protect operators from scatter radiation. Make sure to install the backscatter

shield properly by following the steps before operation.

 

 

a) Check the backscatter shield and

collimator before each operation. Do not operate if any of them is broken.

 

 

b) Keep the backscatter shield on the outer

side of the collimator cone as close as possible to the patient.

 

 

c) Operator stays within the significant

zone of occupancy.

 

 

 

 

 

Never stand in the direct line of the

primary X-ray beam. If patient needs assist, give caregiver personal protective

equipment such as an apron to wear or to stand behind a barrier.

 

 

 

 

 

For reducing radiation exposure to operator

and help ensure stability while taking radiographic images, the use of a tripod

is recommended as possible.

 

 

 

 

 

This system is composed of main body and

accessories as follows.

 

 

Main

 

 

- XR-01 Portable X-Ray

 

 

- Beam limiting Device

 

 

Accessories

 

 

- AC/DC Adaptor

 

 

- Power Cord

 

 

- Hand Switch

 

 

- Carrying Box

 

-

Back Scatter Shield (Optional)